The WGH6431 is a flame photometer for measuring potassium, sodium, and calcium. It determines concentration by monitoring the emission produced when sample atoms are excited in a flame. The instrument is designed for quantitative analysis within the listed element ranges. Up to three selected elements can be shown at the same time. An LCD and menu-based keyboard support direct operation. Automatic correlation calculation helps assess calibration quality. Standard curves can be stored, and single-point correction is available for routine adjustment. Preset flame control and flame-out protection support stable and safe combustion. The RS232 interface connects the instrument to a computer or printer, while the moisture separator helps condition the gas stream.
Specification
Power supply: 220 V, 50 Hz Rated power: 250 W Display: LCD Measured elements: K, Na, and Ca K range: 0-100 µg/mL Na range: 0-100 µg/mL Ca range: 0-1000 µg/mL K detection limit: ≤0.004 mmol/L Na detection limit: ≤0.008 mmol/L Ca detection limit: ≤0.050 mmol/L K linearity error: ≤0.005 mmol/L Na linearity error: ≤0.03 mmol/L Ca linearity error: ≤0.07 mmol/L Short-term stability: Relative maximum indication change ≤3% within 15 s Long-term stability: Relative maximum indication change ≤15% over six measurements at 1 min intervals Repeatability: ≤3% Calibration methods: Piecewise-linear, linear, and curve methods as listed Sample uptake rate: <6 mL/min Communication: RS232 Net weight: 10 kg Product dimensions: 400 × 275 × 460 mm
Features
Three-element measurement Calcium range to 1000 µg/mL Automatic correlation calculation Multi-element LCD display Flame-size preset function Stored calibration curves Flame-out safety device Moisture separator and RS232