This instrument is designed to verify the integrity of gloves used in isolators and similar containment systems. It applies and maintains a controlled test pressure within the glove. Pressure change is monitored and compared with defined acceptance values. The system determines whether the tested glove remains intact or exhibits unacceptable leakage. A large color touchscreen allows test parameters and sample information to be entered. Detailed numerical results and complete pressure curves support evaluation of glove performance. Automatic program selection can match the test sequence to different glove specifications and parameter settings. Electronic signatures and tiered user management support controlled operation and responsibility tracking. Communication software and configurable protocols allow integration with centralized automation systems.
Specification
Power supply: AC 220 V, 50 Hz Rated power: 1000 W Adjustable test pressure: 300 to 1500 Pa Leakage rate: ≤10 Pa Noise level: ≤50 dB(A) Operating mode: Continuous Weight: Approximately 4.5 kg Dimensions: 370 × 220 × 210 mm Display: Large high-resolution color touchscreen Input methods: Chinese, English, numbers, and punctuation Data output: Detailed test data and complete test curves User control: Electronic signature and tiered user management Communication: Host-computer software and customizable protocol External requirement: Clean compressed-gas source
Features