The heater, heat exchanger, circulation pump, valves and pipelines of Techfors-S Cell can be easily observed on the back of the instrument, so that it is convenient to check the working status of the temperature control system at any time
The motor of the Techfors-S Cell adopts the plug-in connection method, and the motor is placed directly on the
* Techfors-S Cell adopts magnetic coupling stirring system, Marine type stirring paddle, suitable for various mammalian cells, insect cells, etc.
* Fully automatic in-situ sterilization: The sterilization of Techfors-S Cell is automatic. During sterilization, steam is passed into the designed double-layer jacket for heating. Steam sterilization is a method of sterilization that ensures sterility. It can be equipped with a compact steam generator, which is fixed at the bottom of the support system, so that in-situ sterilization can be achieved without centralized supply of steam. This design saves a lot of experimental space occupied by traditional steam generators, allows the steam generator to move freely with the entire culture system at the same time, saves a lot of unnecessary pipelines, and avoids the hidden danger of additional pollution to the culture system.
Sterile Sampling Valve
* The sampling valve located at the bottom of the Techfors-S Cell tank, at the center, center and Z-lowest point of the bottom, allows users to discharge the liquid in the tank every time they collect the culture medium or clean it without residue
* The sampling valve has a separate steam sterilization pipeline, which can also be steam sterilized before each sampling to ensure sterility, and the culture solution can be exported through the transfer pipeline, or directly inoculated into the reactor for expanding culture
* In the jacket, the steam path and the water path are separately entered.
* The filter membranes of the air inlet and outlet of the tank are side-mounted, which saves a lot of space on the tank cover and is convenient for users to operate and clean.
* The RSR-232 interface can be monitored in real time by computer software.
More extended functions
* Oxidation-reduction measurement system for determination of oxidation-reduction potential
* External analog peristaltic pump, providing variable rate feed
* Glucose and other analyzers
* Exhaust gas analyzer
* Pressure measurement and control system
* Standard 4+1 (Air+O2+N2+CO2/Air) gas mixing station, optional thermal mass flow meter JQ control mixed gas flow can be applied to most cell culture
* Steam generator etc.
* Refrigeration equipment
* Sterilization equipment - Bunsen burner
* Vacuum filter unit, air compressor air supply unit
* Liquid operating unit
* External balance weighing equipment