The manufacturer of the Dionex Thermostatted Column Compartment TCC-100 is Thermo Fisher Scientific. The Dionex Thermostatted Column Compartment TCC-100 is designed to perform equally well as a dependable system for routine analyses or as a sophisticated research instrument, especially for HPLC applications. The temperature can be set on the instrument itself, using the front panel controls or remotely, via Chromeleon. The design has been optimized for minimum dead volume and maximum efficiency.
Specification
Condition: Pre-owned Environmental Conditions: Range of use: Indoor use Temperature: 10°C to 35°C (50°F to 95°F) Air humidity: 40 to 85% rel. humidity, non-condensing Overvoltage Category: II Polution degree: 2 Operating Conditions: Temperature range: +5°C to +85°C (+41°F to +185°F) When cooling: max. 15°K from ambient Temperature Setting: In 0.1°C increments Temperature Accuracy: ± 0.5°C Temperature Stability: ± 0.1°C Temperature Precision: ± 0.1°C Heat-up/Cool-down Time: Less than 20 minutes from 20°C to 50°C Less than 20 minutes from 50°C to 20°C average change per minute: 2-4°K Column Capacity: 3 columns (maximum length: 30 cm) Control: All functions controllable via LAN or USB Permanent status transfer Program-controlled by Chromeleon or stand-alone operation Switching Valve: Options: 2-position/6-port or 2-position/10-port switching valve Maximum backpressure: 345 bar (5000 psi) Switching time: 110 ms Glp: When controlled by Chromeleon: Dionex AutoQ routines for Instrument Qualification Column identification system for 2 columns
Features
Refurbished to meet manufacturer's specification In excellent condition Thoroughly cleaned Certified with warranty