Determination of pharmaceutical compounds in skin by imaging matrix-assisted laser desorption/ionisation mass spectrometry (CAT#: STEM-ST-0296-LJX)

Introduction

Matrix-assisted laser desorption/ionisation (MALDI) quadrupole time-of-flight mass spectrometry (Q-TOFMS) has been used to detect and image the distribution of a xenobiotic substance in skin. Porcine epidermal tissue was treated with ‘Nizoral’, a medicated shampoo containing ketoconazole ((±)-1-acetyl-4-[p-[[(2R,4S)-2-(2,4-dichlorophenyl)-2-(imidazol-1-ylmethyl)-1,3-dioxolan-4-yl]methoxy]phenyl]piperazine) as active ingredient. Following incubation for 1 h at 37°C all excess formulation was washed from the surface. A cross-section of the drug-treated tissue was then blotted onto a cellulose membrane, precoated in matrix (α-cyano-4-hydroxycinnamic acid (CHCA)), by airspray deposition. In separate experiments the tissue surface was treated with Nizoral within a triangular former, and subsequently blotted onto a matrix-coated membrane. Sample membranes were then mounted into the recess of specialised MALDI targets with adhesive tape. All samples were analysed by MALDI-TOFMS using an Applied Biosystem ‘Q-star Pulsar i’ hybrid Q-TOF mass spectrometer fitted with an orthagonal MALDI ion source and imaging software. Detection of the protonated molecule was readily achievable by this technique.




Principle

In a very small area and a very short time interval (ns order of magnitude), the laser delivers high-intensity pulse energy to the sample under test, causing it to desorption and ionize instantaneously without thermal decomposition. MALDI is a mass spectrometry ionization method for direct evaporation and ionization of non-volatile samples.

Applications

For measuring the molecular weight of biological macromolecules, such as the molecular weight distribution of peptides, proteins, nucleic acids, polymers and oligomer analysis.

Procedure

1. Mix the sample with the appropriate matrix material and load it onto the metal plate.
2. Pulsed laser light is used to irradiate the sample and trigger ablation and desorption of the sample and matrix materials.
3. Analyte molecules are ionized by protonation or deprotonation in the thermal plume of the ablated gas and are then accelerated to a mass analyzer for analysis.

Materials

• Sample Type:
Porcine epidermal tissue

Notes

1. During shutdown, if the nitrogen is not turned off, the pressure should be properly lowered to avoid moisture.
2. Keep an eye on instrument drift during manual measurement. If there is drift, the instrument needs to be calibrated with standard peptide or standard protein.